DEAE-60 fraction analysis.
NS 1500 fraction analysis.
Comparison of the Macherey-Nagel® DEAE-60 to Hydrocell® NS 1500 columns for Phosphorothioate Oligonucleotide [ps-(pT)11T] separations.
Volatile buffers save time for desalting fractions for other studies.
Conditions:
Buffer A: 25 mM triethylamine bicarbonate (~0.35% v/v), 20% MeCN, pH
6.4.
Buffer B: 1M triethylamine bicarbonate (~14% v/v), 20% MeCN, pH 7.6.
Gradient: 1 mL/min., 1%/min., 0 - 100% B.
Sample:
Crude 12mer Thymidine Phosphorothioate Oligonucleotide [ps-(pT)11T].
Note: Failure sequences are not all baseline resolved. The shoulder
(35.73 min.) is possibly an isomer of ps-(Tp)11T,
since it has the same charge as the 37.1 min.peak and MALDI identification
has given conflicting results.
- Hydrocell NS 1500 Part Numbers and Prices
- MN Part Numbers and Prices